Bolus Wizard feature settings
Information Setting
Active insulin time:
The Bolus Wizard feature uses this time to calculate the active
insulin in your system (see the About active insulin section in
this chapter). Use your healthcare professional's recommenda-
tion for the active insulin time that best represents the insulin
type you use and your physiological insulin absorption rate.
range: 2-8 hours
Number of Hours: _____
How the Bolus Wizard feature works
1
If you want your current blood glucose to be factored in, enter your BG Reading.
automatically from the meter (refer to the Meter option section in this chapter), or
manually by selecting the
button.
2
If you are going to eat, enter your food amount in grams or exchanges.
3
The Bolus Wizard feature will calculate a bolus for you. An ESTIMATE DETAILS screen will appear with
your estimated total bolus amount.
Bolus Wizard warnings
When using the Bolus Wizard feature, the pump may display a HIGH BG, LOW BG, and MAX BOLUS EXCEEDED
warnings.
HIGH BG
If your blood glucose in the ENTER BG screen is above 250 mg/dL (13.9 mmol/L), the Bolus Wizard feature
displays a HIGH BG warning. Read the instructions, then press ACT or ESC to clear the message. You can then
continue programming to deliver your bolus.
LOW BG
If your blood glucose in the ENTER BG screen is below 70 mg/dL (3.9 mmol/L), the Bolus Wizard feature
displays a LOW BG warning. Read the instructions, then press ACT or ESC to clear the message. You can then
continue programming to deliver your bolus.
Using the Bolus Wizard feature