Bolus Wizard feature default settings
Item Default Set-
ting
Limits Increments Warning Limits
carb units: grams none none none
ins to carb (or
exch) ratio:
15 grams/u
or
1 unit/exch
1 - 200 g/u or
0.075 - 15.0 u/exch
1 g/u or
0.1 u/exch1 g/u
5 - 50 g/u or
0.3 - 3.0 u/exch
(insulin) sensi-
tivity:
2.8 mmol/L
or
50 mg/dL
10 - 400 mg/dL or
0.5 - 22.2 mmol/L
1 mg/dL or
0.1 mmol/L
20 - 100 mg/dL
or
1.1 - 5.6 mmol/L
BG Target: 5.6–5.6
mmol/L
or
100–100
mg/dL
60 - 250 mg/dL or
3.3 - 13.9 mmol/L
1 mg/dL or
0.1 mmol/L
90 - 140 mg/dL
or
5.0 - 7.8 mmol/L
Active Ins Time: 6 hours 2 - 8 hours 1 hour None
Pump specifications
196