Within 20 seconds, the green light on the transmitter will flash for about 10 seconds when properly
connected. The Sensor feature must be turned On on your pump. Go to the RECONNECT OLD SENSOR
screen to start your sensor.
Main > Sensor > Sensor Start > Reconnect Old Sensor
You should see the sensor icon
on the pump screen, which means that the transmitter and the sensor
are communicating.
4
Within the next seven to eight minutes, go to the SENSOR STATUS screen to look for the Sensor ISIG value:
a. The sensor ISIG value on this screen should be between 24.00 - 29.00 nA. This range of ISIG values
means that the transmitter electronics are working properly. Since the transmitter is sending signals
correctly, the sensor must have caused the alert. Remove and discard the sensor. Insert a new sensor
in a new site.
b. If you see a Sensor ISIG value that is less than 24.00 nA or more than 29.00 nA, contact the HelpLine.
It may be time to replace your transmitter.
Disconnecting the tester
1
Hold the transmitter body as shown and pinch the side arms of the tester.
2
With the tester arms pinched, gently pull the transmitter away from the tester. To save transmitter
battery life, do not leave the tester connected after testing.
Troubleshooting and alarms
171