Accuracy of Guardian RT readings
In this study, YSI measurements (taken every half hour) were paired with the corresponding Guardian RT
reading (taken every five minutes). Pairing was done by selecting the Guardian RT value closest in time to
the YSI test result. Agreement was analyzed by comparing paired glucose measurements.
Agreement between the matched pair was estimated by evaluating the difference between the Guardian RT
reading and the YSI measurement. The difference between them was calculated as a percentage of the YSI
(Mean Absolute Percent Difference). The bias was also calculated, and it is defined as the overall difference
between the Guardian RT glucose values and the YSI values. The paired glucose measurements are
summarized below.
Number of Paired Glucose Measurements
3941
Mean Absolute Percent Difference (± SD) 19.7 ± 18.4%
Bias -0.8 mmol/L (-15.0 mg/dL)
The accuracy of the Guardian RT was also evaluated by calculating the percentage of Guardian RT readings
within 20 percent and within 30 percent of the YSI reading (or within 1.1 mmol/L (20 mg/dL) in the low
glucose range). Results are shown below.
Plasma Glucose
Range (mg/dl)
Plasma Glucose
Range (mmol/l)
Number of Paired
Readings
Percent Within
20%
Percent Within
30%
Overall 3941 62% 79%
40-80
a
2.2-4.4 356 68% 68%
>80-120 >4.4-6.7 769 60% 77%
>120-240 >6.7-13.3 2362 62% 81%
>240 >13.3 454 61% 82%
a. For the Low glucose range, 2.2-4.4 mmol/L (40-80 mg/dL), the value shown is the percent within
1.1 mmol/L (20 mg/dL).
Sensor accuracy
209
Appendix A