Connecting the transmitter to the sensor
Before connecting the transmitter to the sensor, you must have the transmitter fully charged and set up (see
Charging the transmitter and Entering the transmitter ID sections in this chapter). You must also have the
features set up and the sensor inserted (see the instructions in this chapter and those provided with your
sensor). Allow at least one minute after disconnecting the transmitter from the charger before you can
connect it to the sensor.
1
ಘ
After the sensor is inserted, wait five minutes before connecting the transmitter. Check for bleeding.
Make sure that any bleeding has stopped, then connect the transmitter to the sensor.
2
ಘ
Touch the inserted sensor at back of assembly to prevent movement.
sensor
3
ಘ
Hold the transmitter as shown to line up the two notches on both sides with the flexible side arms of the
sensor. The transmitter's flat side with the label should face the skin.
4
ಘ
Slide the transmitter onto the sensor and push it in firmly until the flexible side arms of the sensor click
into the notches on both sides of the transmitter. In the next 20 seconds the transmitter light will flash
green for about 10 seconds with a proper connection.
5
ಘ
If the transmitter light does not flash, disconnect it from the sensor, wait for one minute and then
reconnect. If the transmitter light still does not flash, charge the transmitter.
6
ಘ
After the transmitter light flashes green, use your pump to start communicating with the sensor and to
start the sensor initialization. See the next section.
Sensor features
125
ಘChapter 8