Normal mode
Mode for standard pump operations for normal basal and bolus delivery.
No special features are active (such as basal patterns, temp basal, and
others). No alarms and no alerts exist.
Special mode
Indicates a special feature is active or an alert condition(s) exists.
Special mode does not restrict any of the pump functions. When the
pump is in Special mode, an open circle appears at the top of the screen
and it beeps or vibrates periodically to remind you of the condition. The
conditions and features that put the pump in Special mode are:
Low Reservoir condition
Low Battery condition
Block feature is on
Dual or Square Wave bolus delivery
Basal pattern A or B is active
Sensor alerts
Temporary basal delivery
Attention mode
WARNING: When the pump is in Attention mode, showing a solid circle icon, all insulin deliveries are
stopped.
Indicates insulin delivery has stopped. This can mean that the pump is
in Suspend mode. It can also mean an alarm is active or an alarm
condition exists that needs immediate attention for insulin delivery to
resume. A solid circle appears at the top of the screen and the pump
will beep periodically until either the pump is taken out of Suspend
mode or the condition is cleared. The screen will show text describing the condition that put the pump in
Attention mode. For example, if the reservoir is empty, EMPTY RESERVOIR will appear on the screen.
The basics
34