Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 522/722/522K/722K pump system (consisting of the MMT-522/722/522K/722K pump and
MMT-7703 transmitter) is intended for use in the electromagnetic environment specified below. The
customer or the user of the Paradigm 522/722/522K/722K pump system should make sure that it is
used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Environment
- Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
Not applicable
±8 kV air
±30 kV air
(<5% relative hu-
midity)
The Paradigm
522/722/522K/722K pump sys-
tem should not be affected by
electrostatic discharge that
might occur under normal con-
ditions of use.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
Not applicable
Not applicable
Not applicable
Not applicable
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not applicable
Not applicable
Not applicable
Not applicable
Voltage dips, short inter-
ruptions and voltage var-
iations on power supply
lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycle
<5% U
T
(>95% dip in U
T
)
for 5 seconds
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Pump specifications
199
Chapter 15