Insulin delivery default settings
Menu Item Default Set-
ting
Limits Increments
Bolus Menu: Bolus Wizard feature: Off
Easy bolus: On
Easy bolus step: 0.1 u 2 u
Dual/Square bolus: Off
Maximum bolus: 10.0 u 0 - 25 u (per sin-
gle bolus)
BG Reminder: Off 0:00 - 5:00 0:30
Basal Menu: Patterns: Off
Maximum basal rate: 2.0 u/h 0.00 - 35.00 u/h 0.05 u
Basal rate: 0.0 u/h 0.05 u
Temp basal type: U/H Max basal rate
setting
0.05 u/h (or 1%)
Utilities Menu: Lock Keypad: Off
(Alarm) History: (no defaults)
Alert type: audio, beep-
med
Auto-off: Off
Low reservoir warning: (20) insulin
units
if units: 1st @ 5 -
50 u; 2nd @ half
amount remain-
ing (if time: 1st
@ 2:00 - 24:00;
2nd @ 1:00 hour
before empty)
20 u
(0:30)
(Time/Date) Time: 12 a.m. (mid-
night)
(Time/Date) Date: 1/1/05
Pump specifications
194