3 Enter your new measurement, then press ACT. The BG measurement must be between 20–600 mg/dL
(1.1–33.3 mmol/L). A message asks if you want to save the new measurement that is shown on the
screen.
If the information is correct, select Save BG, then press ACT. Your BG is saved.
If the Bolus Wizard feature is off, the CAPTURE EVENT screen appears.
If the Bolus Wizard is on and the saved BG is below your BG Target range, the BG Saved screen
appears. This screen indicates that your saved BG is below your BG Target set in the Bolus Wizard
feature, and the pump beeps three times. See Setting the BG Targets, on page 71 for more
information.
If the Bolus Wizard is on, the saved BG is above your BG Target range, and the calculated
correction estimate is more than the preset scroll rate increment, the BG Saved screen appears.
This screen indicates that your saved BG is above your BG Target set in the Bolus Wizard feature,
and the pump beeps three times. Go to the ENTER BG screen of the Bolus Wizard feature to enter
a correction bolus. See Setting the BG Targets, on page 71 for more information.
If you want to use the new BG measurement for calibration, select Save and Calibrate. This
option is available when your sensor is communicating with the pump, and the BG measurement is
between 40 mg/dL (2 mmol/L) and 400 mg/dL (22.2 mmol/L).
If the Bolus Wizard is on and the saved BG is below your BG Target range, the BG Saved screen
appears. This screen indicates that your saved BG is below your BG Target set in the Bolus Wizard
feature, and the pump beeps three times. See Setting the BG Targets, on page 71 for more
information.
If the Bolus Wizard is on, the saved BG is above your BG Target range, and the calculated
correction estimate is more than the preset scroll rate increment, the BG Saved screen appears.
This screen indicates that your saved BG is above your BG Target set in the Bolus Wizard feature,
and the pump beeps three times. Go to the ENTER BG screen of the Bolus Wizard feature to enter
a correction bolus. See Setting the BG Targets, on page 71 for more information.
If the information is not correct, select Cancel, then press ACT. The CAPTURE EVENT screen
reappears with Enter BG selected. Repeat the procedure and enter the correct information.
Entering insulin injection information
Follow these steps to enter the amount of insulin you use.
1 Go to the ENTER INSULIN screen.
Main > Capture Event > Insulin Marker
The ENTER INSULIN screen flashes with dashes or the insulin amount that you most recently entered.
Basic programming 39