BOLUS STOPPED
If this alarm occurs, it is very important to check bolus history to review how much of the bolus was
actually delivered. If necessary, reprogram a bolus with the amount not yet delivered.
BUTTON ERROR
Occurs if a button has been continually pressed for more than three minutes.
CHECK SETTINGS
When this alarm is active, you should check and/or reprogram your pump settings, including the time/
date.
E (ERROR)
An E-error alarm will show the letter E followed by two numbers. E-alarms cause all insulin delivery to
stop, the pump resets, and all your settings are cleared. After receiving this alarm, note the error number
then call our 24 Hour HelpLine for assistance.
EMPTY RESERVOIR
There is no insulin in the reservoir. Change the reservoir immediately.
FAILED BATT TEST
The pump tests the voltage of each battery installed. This test ensures a battery with low voltage is not
used. If the battery does not have enough voltage, this alarm will occur. The pump will not function and
the battery must be replaced. (Always make sure that you install a NEW battery into the pump.)
FINISH LOADING
You have not completed filling the infusion set with insulin. Clear the alarm. This resumes basal delivery.
See the Filling the cannula section in the Starting on insulin chapter to fill the infusion set cannula with
insulin. If your infusion set has a needle cannula, press ESC to skip this step.
Troubleshooting, alerts and alarms166