CAL ERROR
Reason: An error occurred when entering a new meter BG measurement to calibrate the system. Some
possible causes are:
An incorrect blood glucose number was entered from the meter into the pump.
The entered blood glucose measurement was not current.
Your blood glucose is rising or falling rapidly.
The sensor needs more time to stabilize after being inserted.
The sensor is no longer reading the sensor glucose correctly.
Action:Follow the guidelines in the Calibrating the sensor section in the Sensor features chapter to enter
a new meter BG for calibration. If you receive a Cal Error on your second calibration, a CHANGE SENSOR
alert will occur. Call our 24 Hour HelpLine if you have questions.
METER BG NOW
Reason: A meter BG measurement is needed immediately to calibrate your sensor so that you can
continue receiving sensor glucose readings.
Action: Take and enter a meter blood glucose measurement. Follow the guidelines in the Calibrating the
sensor section in the Sensor features chapter.
METER BG BY
Reason: A meter BG measurement must be entered by the time that is shown to calibrate your sensor. The
METER BG BY alert is also known as the CAL REMINDER alert.
Action: Enter a blood glucose measurement to avoid a METER BG NOW alert.
LOW XX MG/DL (XX = SG measurement)
Reason: The sensor glucose number is equal to or lower than your low glucose limit. If you do not set a
Low Glucose Limit and do not set the glucose alerts ON, then you will not get a Low sensor glucose alert.
The pump plays four consecutive tones, in falling pitch, if an audible beep has been selected as the alert
type.
Troubleshooting, alerts and alarms162