Easy bolus setup
1 Go to the EASY BOLUS OPTION screen.
Main > Bolus > Bolus Setup > Easy Bolus
Select On/Set, then press ACT.
NOTE: If you are using the remote control, the Easy Bolus must be set to on.
Step value setup
You can set the step value from 0.1 to 2.0 units, or to the set maximum bolus limit of less than 2.0 units
(factory setting: 0.1). Set the step to a number that is convenient to use and easy to multiply.
Example:
2.0 unit Easy bolus with a step
amount set at 0.5 units
Total # Steps = 4
Total # Button Presses = 4
1 Step = 0.5 units
1 Step = 0.5 units
1 Step = 0.5 units
1 Step = 0.5 units
4
3
2
1
0
0 0.5 1.0 1.5 2
Total # Units
1 The EASY BOLUS ENTRY screen flashes the step value. Change the value, then press ACT. The step
value is the increment you will use for your Easy bolus.
2
The screen will return to the BOLUS MENU. Your step amount is now programmed and Easy Bolus is
ready to use. Exit the menus.
Optimizing pump therapy88