Setting the High Repeat
After you receive and clear a High Glucose, Rise Rate of Change, or High Predictive alert, the alert will
repeat until the condition that has caused the alert is resolved. The High Repeat feature allows you to set
how frequently you want the alert to repeat after you clear it the first time. For example, if your blood
glucose levels usually take one to two hours to fall after a correction bolus, you may want to set the High
Repeat to one or two hours to avoid unnecessary alerts.
Example
You set your High Repeat to 15 minutes. You receive an alert at 1:00 PM and immediately clear it.
However, the condition that caused the alert still exists, so you will receive a second alert at 1:15 PM,
15 minutes after you cleared the first alert. If you immediately clear the alert again, then the alert will
repeat at 1:30 PM and continue to repeat every 15 minutes until you resolve the condition that caused the
alert. (This example also applies to the Low Repeat.)
To set up your High Repeat:
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Hi Repeat, then press ACT. The SET HI ALERT REPEAT screen flashes the default High Repeat
time of 1 hour (1:00).
3 Select your High Repeat time. The time must be between 5 minutes (0:05) and 3 hours (3:00).
4 Press ACT. The EDIT SETTINGS screen shows the High Repeat value set.
You are now ready to set up your Low Repeat.
Setting the Low Repeat
After you receive and clear a Low Glucose, Fall Rate of Change, or Low Predictive alert, the alert will
repeat until the condition that caused the alert is resolved. The Low Repeat feature allows you to set how
frequently you want the above alerts or alarm to repeat after you clear it the first time. The Low Repeat
feature works similar to the High Repeat one. See Setting the High Repeat section in this chapter for an
example.
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Low Repeat, then press ACT. The SET LO ALERT REPEAT screen flashes the default Lo Alert
Repeat time of 20 minutes (0:20).