LOW BG
If your blood glucose in the
ENTER BG screen is below 70 mg/dL (3.9 mmol/L), the Bolus Wizard feature
displays a LOW BG warning. Read the instructions, then press ACT or ESC to clear the message. You can
then continue programming to deliver your bolus.
MAX BOLUS EXCEEDED
Bolus Wizard feature will not deliver more than the limit set for your maximum bolus. If the Bolus Wizard
feature calculates a bolus amount that is larger than your max bolus limit setting, the message, MAX
BOLUS EXCEEDED will appear. If this happens, do these steps:
1 In the MAX BOLUS EXCEEDED screen, press ACT to continue your bolus programming. The
EST : MAX
screen appears with the estimated and maximum bolus amounts. Continue to the next step.
If you do not want to continue, press ESC to cancel and the screen will return to the ENTER BG screen.
2 In the EST : MAX screen, press ACT again to continue your bolus programming.
3 The SET BOLUS screen appears with the maximum bolus amount flashing. Enter the bolus amount. This
amount cannot be more than the max bolus amount. Press ACT.
4 The BOLUS DELIVERY screen appears showing the insulin units being delivered.
5 The pump will beep/vibrate after it has completed the insulin delivery.
How to program the Bolus Wizard feature
Y
ou need your personal settings from the Bolus Wizard feature settings table to setup the Bolus Wizard
feature. Your Bolus Wizard feature settings are programmed in the EDIT SETTINGS screen.
Main > Bolus > Bolus Wizard Setup > Edit Settings
Once the settings are programmed, you do not have to program them again unless the values change.
After you program one setting, the screen will automatically go to the next required setting. After you
program all your settings, review them as described in this section to make sure they are set correctly.
Instructions for programming the Bolus Wizard feature settings are in the next paragraphs. Program your
settings in the order described to make sure you program all the settings. If you have not completed
setting up all of the required settings, MISSING INFO screen appears. It lists the required settings for this
feature. You must program the listed settings before you can use the Bolus Wizard feature.
Using the Bolus Wizard feature76