
2 Push the tester into the transmitter until the flexible side arms of the tester click into the notches on
both sides of the transmitter
.
3 Within 20 seconds, the green light on the transmitter will flash for about 10 seconds when properly
connected. The Sensor feature must be turned On on your pump. Go to the RECONNECT OLD SENSOR
screen to start your sensor.
Main > Sensor > Sensor Start > Reconnect Old Sensor
You should see the sensor icon
on the pump screen, which means that the transmitter and the
sensor are communicating.
4 Within the next seven to eight minutes, go to the SENSOR STATUS screen to look for the Sensor ISIG
value:
a. The sensor ISIG value on this screen should be between 24.00 - 29.00 nA. This range of ISIG values
means that the transmitter electronics are working properly
. Since the transmitter is sending
signals correctly, the sensor must have caused the alert. Remove and discard the sensor. Insert a
new sensor in a new site.
b. If you see a Sensor ISIG value that is less than 24.00 nA or more than 29.00 nA, contact the
HelpLine. It may be time to replace your transmitter.
Disconnecting the tester
1 Hold the transmitter body as shown and pinch the side arms of the tester.
Troubleshooting and alarms168
Komentáře k této Příručce